I just got a Picoscope (the 6402C) and i am trying to use it with Labview. I've tried running the examples that are given with the drivers but i've come across some problems.
First i can't run the program, a indow appear signaling "error 5003" "Pico not found" but the picoscope is connected and the light is switched on so I dont know what the problem is. Have you got any ideas please?
The second error that occurs is the 5012 "Pico invalid Handle". I dont really know what it means, so i'll be haapy if someone could give me some insights.
Which version of LabVIEW are you using? Is it 32-bit or 64-bit?
The error code 5003 indicates that a device was not found. Please check that the PicoScope 6 software is not running with a connection to the oscilloscope.
Try another USB port and ensure that the power supply is connected to the PicoScope.
The 'invalid handle' error is caused by calling a function when a device is not connected i.e. a valid handle value has not been passed to the function call.
You can add extra channels by adding controls to the relevant ports on the PS6000 Settings sub-vi block in the Block diagram. You will then need to adapt the rest of the example accordingly.
The programm given in the example doesn't allow us to record the data in a file so I thought about using the function "Write in a measurement file". In doing so i discovered that i couldn't record my data in psdata format. Have you got some ideas about what i could do to record my data in a file?
You can set up a logic trigger e.g. trigger when the conditions for channel A AND channel B are met at the same time, or when the condition for channel A OR channel C is true but it is not possible to set up independent triggers for two channels.
I want to set a trigger if the conditions in A and C are met. So, do I have to use a "Case structure" or just add another "Trigger settings" and use a Boolean "AND"?
In the PicoScope6000.llb file there is a sub-vi for Advanced Triggering. Place this block in your LabVIEW block diagram after the settings sub-vi and you should be able to configure the trigger as required (add controls to the inputs of this new sub-vi as required).