I've just received the TH-03 Themperature Data Logger and installed the PicoLog software and it seems to work Ok. However I like to used the Labview drivers which seems to be available for only old Labview versions like 4,5,6.
Currently I use the files found in the library TH03.LLB in the software development kit R1.0.0
When I open It in Labview 8.5.1 it says the the files are to Old to be converted. Labview 7.1 manage to open the files but with a broken link in the Call Library Function Node of the th03_channel.vi I fixed this broken link by reconfiguring the link to the th0332.dll by choosing th03_labview under function name.
The vi is now able to run. I’ve connected the TH03 to COM 1. And therefore choose port 1 in the front panels controller. However I get no response from the VI. The ok and the temperature indicator are 0. Is there a newer labview driver available or is there another working fix for these labview files?
We are planning to buy the TH-03 Thermistor Converter. I downloaded the TH03.llb above and open the example. It gave an error:
Call Library Function Node 'th0332.dll:_th03_labview@20': Library not found or failed.
The library specified for this node cannot be found or cannot be loaded. Right-click the Call Library Function node and select Configure, then choose the correct library name or path.
How can I solve this?
Thanks!
Last edited by luisbgomez on Wed Sep 22, 2010 12:15 pm, edited 1 time in total.
Robin wrote:
Drivers/examples are included in the siftware development kits (SDKs)
Robin
Thanks Robin,
I downloaded the kit, but include an earlier version I cannot open with LV8.6. I could try to open the llb with an earlier version, I could install on an old computer, but it seems from earlier posts there is an updated version already available.
I opened it and re-saved it in version 8.2. You should be able to open it now.
I noticed that you changed your previous post. Regarding the library, make sure that the dll is in the same folder as the llb file. If it still can't find it, double click on the call library function node and select the dll.
Robin wrote:I noticed that you changed your previous post. Regarding the library, make sure that the dll is in the same folder as the llb file. If it still can't find it, double click on the call library function node and select the dll.
Robin
Yes, I cahnged the previous post after seen in that you have included the zip file.
But still can't find the dll. is it included in the zip file? perhaps I don't have rights to see dll files from from my work computer
The TH03 is an old product and no longer supported, the previous post on this thread was from 5 years ago.
I would suggest just talking directly to the device over the serial port using LabVIEW serial communication. The protocol is defined on page 14 of the manual
From following the manual Instructions I still cannot seem to get it working. The issue seems to lie with the version of labview being too new. Having gone through the errors in labview I am at the point where I need to call the library fucntion. I select the th0332.dll and it says "the library selected is not valid for the current platform. For example, you are running 64-bit labview and the library might be 32-bit"
Is there a way around this? Such as an updated dll file. Or is just that the library is too old to communicate with the newer version of labview?