An additional 2.5 MHz setting would indeed not only solve my problem, but be appreciated from anyone using this DSO as FFT spectrum analyzer as serious lab tool. I do, and selected it because there is no competition for what the hardware can do. I don't use the DSO part at all. Martyn, how long you ...
Well, I really don't know what to say. Is this a serious measurement instrument costing 1,200 €? Or a toy for my child to play? What else did you 'round up' in this software to make it look nicer? Is there anything I can still trust, or is this an example for the accuracy I have to expect from Picot...
So, for example, if you select a bandwidth of 2MHz for the PicScope 4262, the sample rate used is the nearest sample rate available that would be 2 times the Bandwidth which is 3.333MS/s (the only sample rates available are based on the maximum sample rate for the PicoScope model divided by factors...
hexamer - thanks for coding this nice tool and providing it to the community! I checked all log entries and got this: Status: 4262 S/N: CU988/013 successfully initialized. Status: Starting frequency step 1 (1.000 Hz) Status: Starting frequency step 1, range try 1 Status: Setting signal generator fre...
4262 with latest Picoscope 6.12.7. Choosing a spectral range of 2 MHz the x axis is labelled 1.6 MHz, with a max of 1.65 MHz to the right. I'm a new user but heard that in older versions there was no 2 MHz option at all, but a 1.7 MHz. So it looks like this choice is wrong and should be 1.7 MHz?
I really would like to try out this software/tool - unfortunately I can't get it working. Windows 10 64, not the latest Creator's update, FRA 0.6. I have a 4262, which is detected on start of FRA. All standard default settings. I hit go, the log says it will start with the first frequency, I hear a ...