Hi,
I have an EL005 + EL041 connected to a sensor with a 4...20mA output. I connected the 4..20mA with a 2.2ohm shunt resistor to an input of the EL041, so I have a max signal of about 44mV. The system works, but I would like to ask a couple of questions:
1. is it possible to increase the sampling speed? Now it is 1 reading/minute, I need 1 reading per second or at least 1 every 10 second
2. could I scale the output of the sensor so I can see the right engineering units instead of the mV?
The EL041 does not support scaling, however the EL037 converter does and will enable you to connect a 4-20mA signal.
If you want to increase the sampling speed however, you will need an entirely different system. The Enviromon system cannot sample faster than once per minute so you would need a different data logger and Picolog software. This software also supports scaling.
I recommend looking at the ADC-20 data logger as this will suit your input and sampling needs.
Just found out that EL005 fastest sampling rate is once per minute.
I need faster sample rate for recording door switch open/close state which range fall in range between 20 - 40 seconds.
Kindly advice product suitable for that purpose.
Thanks for the information, hope to get more detail explanation.
I learnt that ADC11 have no power supply, neither battery since here powers is frequently blackout, it will be much better if there is a battery powered and own memory feature.
If available please advice complete list part no. including logger/sensor/switch/cable that fit my need ie. fast sample rate, high sensitivy, battery backup power, own memory and PC networking something like mix of EL005 + ADC11