I have a question about the pl1000 SDK :

In the examples they are using :
status = pl1000SetInterval(g_handle, &usForBlock, nSamplesPerChannel, channels, nChannels);
status = pl1000Run(g_handle, nSamples, BM_STREAM);

where nSamples = nSamplesperChannel * nChannels

However in the API description is written for both pl1000SetInterval and pl1000Run :
the total number of samples to be collected across all channels, so both should use nSamples.

What is correct ?
Posts: 0
Joined: Mon Nov 20, 2017 2:45 pm

Re: pl1000SetInterval : sdk example != api description

Postby Hitesh » Tue Dec 05, 2017 5:25 pm

Hi fred,

Unfortunately, there are some errors in the Programmer's Guide and the examples which will be corrected.

Please use the number of samples PER channel in the calls to pl1000SetInterval() and pl1000Run().