A3076 750 MHz active oscilloscope probe

The Pico A3076 Series high-impedance active oscilloscope probe has minimal impact on the signal being probed with maximal signal transfer to the PicoScope 6000E Series. Its ergonomic design allows comfortable handheld use with a button to start and pause capturing in PicoScope 6.

The Intelligent Probe Interface powers the probe from the scope and facilitates automatic probe detection and unit scaling. The probe can set scope attenuation and automatically switch to 50 Ω input impedance.

With an input resistance of 1 MΩ and capacitance of 0.9 pF, this active probe offers high input impedance up to 750 MHz. These characteristics make this versatile probe ideal for many of your day-to-day measurements

A3000 Series active oscilloscope probe kit
In stock. Available for despatch.
£879
Title Language Issue Size Updated All products All languages All documents Status
A3000 Series Active Oscilloscope Probe Quick Start Guide English
Français
Deutsch
Italiano
Español
中文 (简体)
한국어
日本語
1 5 MB Sep 24 2021 english
french
german
italian
spanish
chinese-simplified
korean
japanese
open
A3076 and A3136 Active Single-Ended Probes EU Declaration of Conformity English 2 434 KB Nov 15 2022 english open
A3000 Series Active Oscilloscope Probe User’s Guide English 1 1 MB Nov 26 2021 english open
A3000 Series Active Oscilloscope Probe Data Sheet English 1 1 MB Sep 24 2021 english open
A3000 系列有源示波器探针数据表 中文 (简体) 1 1 MB Nov 16 2021 chinese-simplified open
A3000 시리즈 액티브 오실로스코프 프로브 데이터 시트 한국어 1 1 MB Dec 07 2021 korean open
A3000シリーズアクティブオシロスコーププローブデータシート 日本語 1 1 MB Nov 16 2021 japanese open
Datenblatt aktive Oszilloskoptastköpfe der Serie A3000 Deutsch 1 1 MB Nov 16 2021 german open
Fiche technique des sondes d’oscilloscope actives de série A3000 Français 1 1 MB Nov 16 2021 french open
Hoja de datos de las sondas para osciloscopios activos serie A3000 Español 1 1 MB Nov 16 2021 spanish open
Scheda tecnica delle sonde per oscilloscopio attive serie A3000 Italiano 1 1 MB Nov 16 2021 italian open
A3000 Series Active Oscilloscope Probe UK Declaration of Conformity English 1 430 KB Feb 23 2022 english open

Accessories for the A3076 750 MHz active oscilloscope probe

Product Price Quantity
A3000 replacement consumables kit A3000 Series replacement consumables kit
PQ275
£17
A3076 active probe specifications
Probe bandwidth DC to 750 MHz into 50 Ω load
System bandwidth DC to 750 MHz (nominal) system bandwidth with PicoScope 6000E Series 750 MHz oscilloscope
Input dynamic range ±5 V (DC + AC peak)
DC gain accuracy ±3% of signal
DC gain accuracy (system) ±4% of signal (nominal) with PicoScope 6000E Series oscilloscope
DC offset accuracy ±3 mV
DC offset accuracy (system) ±(1% of full scale + 4 mV) (nominal) with PicoScope 6000E Series oscilloscope. Offset accuracy can be improved by using the “zero offset” function in PicoScope 6.
DC offset adjustment range ±10 V
DC offset control accuracy ±1% of offset setting, additional to DC accuracy above
Input characteristics 1 MΩ || 0.9 pF
Attenuation 10:1
Oscilloscope input coupling DC 50 Ω
Dynamic performance (typical)
Harmonic distortion –40 dBc (1%) with 250 MHz 1 V p-p sine
Noise 2.5 mV RMS (nominal) referred to probe input
Bandwidth flatness (+1 dB, –3 dB) from DC to full bandwidth
Low-frequency flatness < ±6% (or ±0.5 dB) from DC to 1 MHz
Propagation delay 5.4 ns (typical)
Input voltage
Measurable voltage window ±15 V (DC + AC peak)
Maximum nondestructive input voltage ±30 V (DC + AC peak) subject to frequency derating (derated as shown in A3000 Series Quick Start Guide).
General
Probe detection Automatic with any oscilloscope with Pico Intelligent Probe Interface (IPI)
Probe button Control start/stop capture in PicoScope 6
Operating temperature range 0 to 50 °C
Temperature range for quoted accuracy 15 to 40 °C after 20-minute warmup (system performance specifications assume that both the probe and oscilloscope are within specified operating temperatures).
Storage temperature range –20 to +60 °C.
Operating humidity range 5% to 80% RH non-condensing
Storage humidity range 5% to 95% RH non-condensing
Altitude range Up to 2000 m
Pollution degree EN 61010 degree 2: “only nonconductive pollution occurs except that occasionally a temporary conductivity caused by condensation is expected”
Dimensions (probe body) 108 mm (L) x 16 mm (H) x 12 mm (W)
Cable length 1.2 m
Input type Sockets accepting 0.635 to 0.940 mm round or 0.64 mm square conductors. 7.12 mm pitch.
Weight 75 g
Safety compliance EN 61010-031:2015
EMC compliance Tested to EN 61326-1:2013 and FCC Part 15 Subpart B
Environmental compliance RoHS, WEEE, REACH

Specifications valid over ambient temperature range 15 to 40°C after 20-minute warmup.