Capture count does'n work

Forum for discussing PicoScope version 6 (non-automotive version)
Post Reply
TSANYU
Newbie
Posts: 0
Joined: Fri Oct 23, 2020 2:48 am

Capture count does'n work

Post by TSANYU »

Hello PicoScope Support,
I want to compare the the high end time of PWM in a period.
But I can't capture all high end in a period.
As the picture in attachments, capture count is only 1.
So value, min, max, average are always the same.
how can I capture all high end and get min,max value?
Attachments
擷取.JPG

Gerry
PICO STAFF
PICO STAFF
Posts: 1145
Joined: Mon Aug 11, 2014 11:14 am

Re: Capture count does'n work

Post by Gerry »

Hi TSANYU,

You need to use repeat trigger with multiple buffers of data collected in order to get statistical values of the data. Could you post a psdata file, so that we can see how you have setup the software.

Regards,

Gerry