Hi everyone,
I'm using a PicoScope 2206B with the SDK (PicoSDK_32_10.7.21.227).
The LabVIEW 20.0.1(32-bit) is running on the Windows 10.
This error has occurred.
An error occured while trying to initialize the unit.
Attempt to disconnect and reconnect the unit to the computer.
The previous topic has posted at 2017, the same error as this topic has occurred at 2021.
Should I still use "the examples on the ps2000a folder" described by above topic ?
Now I'm going to use....
Hi, i'm using 2206B with the example ps2000A like you told here but it dosn't work.
it initialize but after that i'm reciving:
----------
Error API function call Call Library Function Node in PicoScope2000aGetBlock.vi->PicoScope2000aExampleBlock.vi
PICO_STATUS Unknown Error
Error code -3903 Decimal
FFFFF0C1 Hex
----------
using PicoSDK_32_10.7.21.227 and labview 2020 32bit. any idea?
i hear some relay switching inside 2206B so at least some comunication is happening.
yes, in fact another example PicoScope2000aExampleStreaming.vi works (still understanding what it does) but at least it capture some data.
i'll try to update shared folder to see if it start working.
now i'm looking at the programmer guide to see if there is something.
i don't know if i can write here also this issue, i've searched in the forum but not find .
in the file C:\Program Files (x86)\Pico Technology\SDK\inc\PicoStatus.h
i think there are few errors. in the value of PICO_INFO 0x00000013E,0x00000013F,0x000000141,0x000000142 seems there is 1 surplus digit. i was looking to have a VI that give an error with a descritpion from the header file.
thanks for the help.
Hi, i'm still here. is there a way to have direct support? i'm going low on time and need picoscope working ASAP.
1- tried the link provided
2- updated this morning SDK 32bit to PicoSDK_32_10.7.22.241
3- still not able to run PicoScope2000aExampleBlock.vi, PicoScope2000aExampleBlockAdvancedTrigger.vi, PicoScope2000aExampleRapidBlock.vi
4- the problem is in: ps2000aGetValues from ..\Picoscope\picosdk-ni-labview-examples\ps2000a\PicoScope2000a.llb\PicoScope2000aGetBlock.vi
5- the error that returns is 1097 "Call Library Function Node in PicoScope2000aGetBlock.vi->PicoScope2000aExampleBlock.vi"
6- using Win10Pro 64bit
7- already uninstalled and installed SDK but nothing changed
Downsampling controls were set to ‘Aggregate' with downsample ratio of '1’
Causing blocks captures (Block, rapidBlock, rapidBlockbulk) to freeze/lockup.
Maybe caused by driver changes also.
I have updated most the 2000a examples with new defaults values ('none' and ratio of 2) also minor change/fixes- https://github.com/picotech/picosdk-ni- ... er/ps2000a