Do you get the same effect if you try channels B to D ? You will note that as you change from 500mV to 200mV a relay will click inside the unit as the input stage is switched to handle the smaller level signals. This will mean that the overvoltage near 10V signal you have at the input connector is s...
Can you post a psdata file so that we can see exactly what data you have.
It can be possible to create a math channel related to the repeat frequency of the ignition signal, and then using the X-Y Instrument replace the X-axis with this math channel.
That does look strange when changing the voltage range from +/-500mV to +/-200mV, it would be useful to know how high a voltage overrange there is, so changing from the +/-500mV range to +/-20V range would be interesting to see.
The OP was after 5 to 15 PicoScopes which is an unusual request. Most cases for adding multiple scopes are unique so we prefer to advise people in relation to their specific needs which is why we will often ask people to email in to support@picotech.com Can you give more detail on exactly why you ar...
There is a function in the API to set the frequency counter although there are currently no examples to show this PREF0 PREF1 PICO_STATUS PREF2 PREF3 (ps4000aSetFrequencyCounter) ( int16_t handle, PS4000A_CHANNEL channel, int16_t enabled, PS4000A_FREQUENCY_COUNTER_RANGE range, int16_t thresholdMajor...
If you look at the Scope button on the top of the display you will see Samples 125 (1290), PicoScope 7 is only collecting 125 samples, but the softare is using sin(x)/x interpolation to bump this up to 1290 which is why you are seeing the ripple. If you are using the same sampling settings as PicoSc...