Search found 25 matches

by egonotto
Sun May 01, 2016 6:47 am
Forum: USB PC Oscilloscopes
Topic: Oscilloscope Mode and AWG Mode
Replies: 3
Views: 12467

Re: Oscilloscope Mode and AWG Mode

Hello,

i don't have the 2204A but i think this is possible with the 2204A too.

I do a sweep from an low frequency amplifier with gain 10 with the 5243A from 1kHz to 500kHz in 1kHz steps.

Best regards
egonotto
by egonotto
Mon Apr 25, 2016 11:45 pm
Forum: USB PC Oscilloscopes
Topic: spikes in triangle from functiongenerator of Picoscope 4262
Replies: 3
Views: 4297

spikes in triangle from functiongenerator of Picoscope 4262

Hello, since few days i have an used Picoscope 4262. It almost as new beside it has only 3 legs :). I try the build in functiongenerator and see a strange behavior of the triangle waveform. Sometimes there are spikes. Insofar i know, this spikes comes not on sine wave and comes not on arbitrary wave...
by egonotto
Thu Aug 27, 2015 11:25 pm
Forum: PicoScope 6 for Windows
Topic: Problem with FFT
Replies: 8
Views: 8739

Re: Problem with FFT

Hallo bennog, PicoScope 6 has a "Spectrum Mode". This should work. I can choose span from 100 Hz to 100 MHz in 20 steps. It makes no sense, if you can only use the 100 MHz span. The attachment shows a 3 MHz square from the internal generator of the PS 5243A with 100 MHz span. But in this case seems ...
by egonotto
Tue Aug 25, 2015 3:46 am
Forum: PicoScope 6 for Windows
Topic: Problem with FFT
Replies: 8
Views: 8739

Re: Problem with FFT

Hello bennog,

thanks for your replay.
How would be your spectrum if you use 90 kHz from extern generator and the equal adjustment in PicoScope as in your picture.
Wrong
Wrong
Ok
Ok
Best Regards
egonotto
by egonotto
Wed Aug 19, 2015 12:43 pm
Forum: PicoScope 6 for Windows
Topic: Problem with FFT
Replies: 8
Views: 8739

Re: Problem with FFT

Look at the 100kHz square with a sample rate of 199 and 496 kS/sec And you will see why the fft is showing the data you posted. That it works at 500 kS/sec is because the source clock of the generator is the same as the clock source of the scope. If you had used a external function generator you wo...
by egonotto
Tue Aug 18, 2015 2:43 pm
Forum: PicoScope 6 for Windows
Topic: Problem with FFT
Replies: 8
Views: 8739

Problem with FFT

Hello,

The FFT of a 100kHz square seems not always correct. See the pictures
Only "500kHz_8Bit.png" looks correct.

I connect the function generator to input A with a BNC-cable.

I use PicoScope® 6 - PC Oscilloscope software version: 6.10.16.2 with PicoScope 5243A

Best regards
egonotto
by egonotto
Sat Mar 28, 2015 2:51 am
Forum: USB PC Oscilloscopes
Topic: noise of PS 5243A to high
Replies: 2
Views: 4547

Re: noise of PS 5243A to high

Hello Andrew, thanks for your answer. I compare my data with the specification in the datasheet: "Noise (on 50 mV range) 8-bit mode 120 μV RMS • 12-bit mode 110 μV RMS • 14-bit mode 100 μV RMS •15-bit mode 85 μV RMS • 16-bit mode 70 μV RMS". This is valid also for the scopes with 200MHz bandwidth. A...
by egonotto
Sun Mar 15, 2015 3:00 pm
Forum: Getting Started
Topic: Picoscope 6 sees only 32MB on PicoScope 5243A
Replies: 1
Views: 2690

Re: Picoscope 6 sees only 32MB on PicoScope 5243A

Hello,

the full memory one can only used in single shot mode.

I think otherwise the scope use 2 buffer.
One buffer samples and the other buffer transfer to PC.

Best Regards
egonotto
by egonotto
Sun Mar 15, 2015 2:28 pm
Forum: USB PC Oscilloscopes
Topic: noise of PS 5243A to high
Replies: 2
Views: 4547

noise of PS 5243A to high

PS5243A_15_noise.jpg PS5243A_8_noise.jpg Hello, the noise of my PS 5243A 100MHz seems to be to high: Bit AC RMS Peak To Peak 8 .... 263uV _ 1.13mV ___ > 120uV 12 .. 107uV _ 1.04mV 14 .. 115uv _ 1.11mV ____ > 100uV 15 .. 106uV _ 1.02mV ____ > 85uV 16 .. 85uV __ 0.810mV ___ > 70uV I use: PicoScope® 6...
by egonotto
Thu Mar 12, 2015 1:14 am
Forum: Getting Started
Topic: Picoscope 6 sees only 32MB on PicoScope 5243A
Replies: 1
Views: 2690

Picoscope 6 sees only 32MB on PicoScope 5243A

Hallo, i got today a new PicoScope 5243A Hardware Version: 1 Driver Version: 1.1.4.20 Firmware Version: 1.3.3.0 / 1.0.24.0 The PicoScope 5243A should have 64MB. But PicoScope 6.10.11.15 use only max 32MB. Maximum Waveforms is set to 1 Some examples: with only 1 Channel Collection time .. Number of S...