Search found 434 matches

by PeterF
Thu May 22, 2008 1:34 pm
Forum: PicoScope 5
Topic: Restoring Picoscope window
Replies: 2
Views: 4573

Hi,
Do you click on the "Restore Down" box (two overlapping squares) when maximised? That should give you a resizeable window.
Also if you right click on the minimised program in the toolbar, you can select "Restore" which does not force it to "Maximise".
Hope that helps.
PeterF
by PeterF
Mon Apr 28, 2008 3:10 pm
Forum: PicoScope 5
Topic: AWG timing is off on the 2203
Replies: 1
Views: 3892

Hi,
We are on the first release of the software for the AWG of the 2203, 2204 & 2205. We have found a few bugs with the incrementing and setting and fixes will be out soon in PicoScope6.
PeterF.
by PeterF
Mon Mar 17, 2008 3:39 pm
Forum: USB PC Data Loggers
Topic: ADC11 USB Help needed to log measured mA
Replies: 1
Views: 4002

Hi Derek, It is difficult to be sure exactly what is causing your problem without being able to view and measure voltages but I will go through some possibilities. Your amplifier is probably being overloaded because the input lines are being pulled out of the common mode voltage range. There are two...
by PeterF
Tue Mar 11, 2008 3:30 pm
Forum: Pre-Sales Advice
Topic: Fast sensor to measure snow temperature
Replies: 1
Views: 4000

Hi, The difficulty is that you will need a robust probe but with very low thermal mass. A thermocouple or a thermistor will be small enough but you will have to get it packaged into a suitable probe case strong enough to push through the ice to the depth you need. A specialist thermoprobe manufactur...
by PeterF
Tue Jan 22, 2008 11:51 am
Forum: PicoScope 5
Topic: Trigger probs PS3205
Replies: 10
Views: 15224

Hi,
The 3223, 3224, 3423 & 3424 all now have advanced triggering in PicoScope6 (6.0.12.2). This includes variable Hysteresis. The 320x series scopes do not at present but this might change in the future.
PeterF.
by PeterF
Tue Jan 22, 2008 11:28 am
Forum: PicoScope 5
Topic: Error in spectrum window
Replies: 2
Views: 4735

Hi, There will be some variation when the bandwidth is changed. The most accurate readings will be when the measured frequency is just below the upper range of measurement (a full scale reading). At higher bandwidth settings, when the peak is only a fraction of the full bandwidth, accuracy will suff...
by PeterF
Fri Jan 11, 2008 8:28 am
Forum: PicoScope 5
Topic: log view of frequency in spectrum analyzer? and x-y?
Replies: 1
Views: 4979

Hi,
My understanding is that here is no log scale available for the x, (frequency) axis. At present, there is no x-y setting either in PicoScope6 but it should be coming soon. In the meantime, if you need x-y, you will have to use PicoScope5, R5.18.0 which does have x-y.
Regards,
PeterF.
by PeterF
Thu Jan 03, 2008 1:46 pm
Forum: PicoScope 5
Topic: Windows 2003
Replies: 2
Views: 4583

Hi,
I am afraid we do not support Windows 2003 server. Our equipment was never intended to to be used under that software. We do not at present have the resources to support it. Sorry.
PeterF.
by PeterF
Fri Dec 14, 2007 3:22 pm
Forum: PicoScope 5
Topic: Amp Hour meter
Replies: 1
Views: 4206

Hi, Your series current sensing resistor must be between the negative terminal of your battery and the common ground of your system. The scope ground must be to the common ground and the scope input to the other end of the resistor. You have a 12 bit scope so the resolution will be good enough with ...
by PeterF
Fri Dec 14, 2007 2:57 pm
Forum: PicoScope 5
Topic: Phase in Spectrum Analyser
Replies: 2
Views: 5257

Hi, I have managed to trace the history now of what happened to that measurement! It was available in earlier versions of PicoScope5 but was causing confusion with many customers who thought it was measuring phase angle between channels A & B. After several people were caught out, we decided to remo...
by PeterF
Fri Dec 07, 2007 2:31 pm
Forum: Pre-Sales Advice
Topic: Multi Channel Datalogging & 1M limit
Replies: 1
Views: 3760

Hi, In PicoLog, each saved file is limited to 1M samples but once that is reached another file with the same name (but including an incrementing suffix) is automatically started. This can go on indefinately until your hard drive is full! Be aware that if you run 8 channels at rates approaching 100S/...
by PeterF
Fri Dec 07, 2007 2:01 pm
Forum: PicoScope 6 for Windows
Topic: dBV in spectrum mode not correct?
Replies: 2
Views: 8492

Hi,
dBm & dBu will only read the same level at a set impedance of 600ohms. dBV should read -2.218 dB lower than dBu but does not in PicoScope6 at the moment, I will be putting in a bug report. Thanks for letting us know.
PeterF.
by PeterF
Tue Dec 04, 2007 12:06 pm
Forum: Pre-Sales Advice
Topic: Advanced Trigger Modes on 3206?
Replies: 1
Views: 3740

Hi,
Advanced triggering is not yet available for the 320x series in PicoScope6 but it is being worked-on. Please subscribe to our newsletter to get news of when new features like this are being introduced.
PeterF.
by PeterF
Thu Nov 22, 2007 12:55 pm
Forum: PicoScope 5
Topic: Exporting Data
Replies: 1
Views: 3887

.TXT files for scope windows Column 1 is the time, usually in microseconds or nanoseconds. Column 2 is the voltage, in either millivolts or volts. If there is more than one channel, the times and voltages for channel A appear first, then the times and voltages for channel B appear below, and so on. ...
by PeterF
Thu Nov 22, 2007 12:19 pm
Forum: Pre-Sales Advice
Topic: PicoScope 2202
Replies: 1
Views: 4081

All timing in our scopes is directly linked to the crystal oscillator and will be 100ppm accuracy. Check by measuring the frequency of an accurate frequency source using the scope frequency measurement. The accuracy is far better than you can resolve on even the largest screen.