Hi, FYI - we have LabVIEW section on the forum - https://www.picotech.com/support/forum20.html Just had look at your LabVIEW code. We don't have any LabVIEW code examples that find the fastest timebase based on the scopes settings but this should be straight forward. Before calling ps3000GetTimebase...
FYI for anyone having connection issues in PicoLog 6 here the USB permissions command that can be run from the terminal.
You will need to restart the Raspberry Pi for it take affect.
Hi, I have updated the streaming example. Correcting the enum values for rising/falling triggers and also some other changes. And today have done some more testing. Yes do agree the falling trigger does seem to behave incorrectly. Yes the trigger reports at sample 0 if the trigger level is below the...
For Windows PC that have never had PicoScope software installed, silent install command option will fail. And will prompt for user interaction "User Account Control" (UAC) dialog box. For silent install to work it first requires windows software certificates to be installed. Here is a zip file that ...
I have attached images of code examples of how to use our old and current error handler. Current- Using errorhandler2.png Old- Using errorhandler.png You may wish change the code in the current errorhandler - (PicoErrorHandler2.vi), so it writes the error text (function, error details) to a file, ra...
With PicoLog 6 using 'Overwrite after fixed time' setting the oldest data will overwritten first. So the recording will always have the last 24 hour recorded. PicoLog 6, works differently to PicoLog 5 as it stores it data in a internal database and does write the data to file like PicoLog 5. But in ...
Relay clatter will be the channel input attenuation relays, switching. All the channels are on by default, set to a default range. I assume you are calling the ps6000aSetChannelOn/Off functions before the call to ps6000aRunBlock? If you calling ps6000aOpenUnit and ps6000aCloseUnit on each run then t...
Hi, With a 2204A connected in spectrum mode, I have set the range 2MHz and the zoom to 100kHz-900kHz
And increased the Number of Bins to max. of 1048576 giving-
Bin width 1.49 Hz
Bins 4096
Collection time 2.58ms
Screenshot-
Hi, We don't have a version of PicoScope software for the raspberry pi devices. But our datalogging software is support on the raspberry pi under Raspbian 32-bit OS. https://www.picotech.com/downloads/_lightbox/picolog-6-raspbian We also offer drivers to write your own software. See - https://www.pi...
Hi, The issue here is due to passing a LabVIEW array into the our dll functions ps5000aSetDataBuffers() and then expecting the array to update with data from the scope, after calling another function. Is against LabVIEW coding convenient. Also it easy to make this mistake in other programming langua...