The default operation for the onboard memory is a segmented circular buffer where you can have up to 1000 waveforms. You can set the number of buffers in Tools->Preferences->General->Maximum Waveforms To get access to the full buffer for a single capture you need to set your trigger mode to single. ...
I don't believe there is a problem with your import.
Could you put all your driver calls up to this point in a simple console application. If that can reproduce your problem, could you please attach that .cs file to this thread?
How far did you get with the installation? 1. What is your output from lsusb with your device plugged in 2. Can you get usbtest to work as root (sudo) 3. Can you get the example program to work as root (sudo) 4. Have you installed the .rules file 5. Have you added your user to the group pico (or the...
You might find that it is the DDS that is giving you the results. If the sampling rate of the DDS is a multiple of 5 and 10MHz it can do a perfect squarewave for those frequencies. Any other frequencies (not an integer divider of half the dds sampling frequency) close to the samplingrate will have a...
You are correct, it looks like the Dr DAQ is not supported by PicosScope 6. Both prodects should be supported by PicoScope 5 though. I believe the feature "install for other devices" is available for PicoScope 5 as well.
Erik wrote:Could you send an email to support@picotech.com describing the steps you take to make the software crash. Also put the version you are using in the email.
When you are comparing ranges and resolution bear in mind that. Our 50 mV range is the same as: 10 mV/div 390 uV/count (with 8 bit resolution, 2000/3000/5000 series) 24.4 uV/count (with 12 bit resolution 3000/4000 series) Our 20 mV range is the same as: 4 mV/div 156 uV/count (with 8 bit resolution, ...
You could use two channels, one measuring each side of the resistor. You could then use a math-channel to calculate the difference i.e. the voltage drop across the resistor.
You could also scale the math channel to give you the current rather than the voltage drop.
Hi, I believe some of the functions in the example program might be leftovers from a version for the ps3000,ps5000 range. For example the ps4000 does not have a signal generator so those functions are not available for the ps4000. The trigger point within a capture is set by the noOfPreTriggerSample...
Assuming you have set it up to trigger on a rising edge there is no bug. It is clearly triggering on a rising edge. To trigger on the first rising edge in a burst you need to setup a more advanced trigger criteria. You could try Pulse Width Negative Pulse Greater than 200 us That should trigger on t...