Search found 3 matches

by pathammer
Tue Mar 20, 2018 4:16 pm
Forum: USB PC Oscilloscopes
Topic: DC measurement error on 4444
Replies: 6
Views: 6458

Re: DC measurement error on 4444

I forgot to ask, is there a way to set the zero offset programmatically? I couldn't find a function in picoscope-4000-series-a-api-programmers-guide.pdf
by pathammer
Tue Mar 20, 2018 4:12 pm
Forum: USB PC Oscilloscopes
Topic: DC measurement error on 4444
Replies: 6
Views: 6458

Re: DC measurement error on 4444

Thanks Gerry! Setting the 'Zero offset' for my channel got my error down significantly. Thanks for aligning my expectations with the equipment spec.
Respectfully,
Patrick
by pathammer
Mon Mar 19, 2018 6:14 pm
Forum: USB PC Oscilloscopes
Topic: DC measurement error on 4444
Replies: 6
Views: 6458

DC measurement error on 4444

Hi there, I have a PS4444 scope (SN: FS786/0025) and have found significant DC measurement error. My test setup consisted of a BK Precision 4053B Waveform Generator whose output connected to my PS under test and a Keysight 34461A reference DMM. I double checked the readings with a Fluke handheld mul...