Search found 3 matches

by mickinuk
Mon Sep 25, 2017 5:01 am
Forum: Getting Started
Topic: Noice level of 2206B MSO
Replies: 3
Views: 5251

Re: Noice level of 2206B MSO

Just found the 1x bandwidth of the 60MHz probe in the specifications - 15MHz.
by mickinuk
Mon Sep 25, 2017 3:20 am
Forum: Getting Started
Topic: Noice level of 2206B MSO
Replies: 3
Views: 5251

Re: Noice level of 2206B MSO

Thanks for the response. I investigated further. It requires the probe to be in 1x mode, as in 10x mode I only get +-200mV, while in 1x mode I get +-20mV and indeed the noise level is 195uV. Thanks.

I believe 1x mode has a big impact on bandwidth, what's the 1x bandwidth with the standard probes?
by mickinuk
Fri Sep 22, 2017 3:48 pm
Forum: Getting Started
Topic: Noice level of 2206B MSO
Replies: 3
Views: 5251

Noice level of 2206B MSO

Can you verify if the noice level I'm experiencing is correct for the 2206 please. The data sheet says <200uV RMS, but I'm seeing 1.8mV RMS, 11.85mV P-P. This is powered from a Mac laptop (Windows version of software) and is the same with or without power. Results are the same if I have both probes ...