Search found 2 matches

by jpelletier
Tue Dec 01, 2015 8:13 pm
Forum: USB PC Oscilloscopes
Topic: Low sample rate acquisition
Replies: 2
Views: 2954

Re: Low sample rate acquisition

The problem that we had: we used a too fast horizontal rate and the scope had a limit of 32 buffers. After that, the data was overwritten. By using a horizontal rate of 1000s/division, we get about 100 samples/s, which is enough for our purpose. It doesn't even fill one buffer. A buffer is 10 divisi...
by jpelletier
Mon Nov 30, 2015 6:27 pm
Forum: USB PC Oscilloscopes
Topic: Low sample rate acquisition
Replies: 2
Views: 2954

Low sample rate acquisition

Hi,
We need to log 2 to 4 waveforms for 10 minutes. A sample rate of 1 sample/sec. would be enough. Is it possible to do this with a Picoscope 6404D ?
Thanks in advance!