Search found 229 matches

by AndrewA
Wed Aug 21, 2019 9:55 am
Forum: LabVIEW
Topic: PicoScope 4824 PICO_SIGGEN_INVALID_SWEEP_PARAMETERS
Replies: 1
Views: 32

Re: PicoScope 4824 PICO_SIGGEN_INVALID_SWEEP_PARAMETERS

Hi, Does the uncomplied version of the code work? I just tested the 64-bit github example- 64-bit PicoScope4000aExampleAWG.vi And it works fine. There is no code diff...
by AndrewA
Wed Aug 07, 2019 2:15 pm
Forum: USB PC Data Loggers
Topic: PicoLog software keeps crashing (5.22 build XP OS)
Replies: 1
Views: 39

Re: PicoLog software keeps crashing (5.22 build XP OS)

Hi,
Can you try running the last version released for Window XP (SP3)-
PicoLog_r5_24_8
by AndrewA
Thu Aug 01, 2019 2:02 pm
Forum: Getting Started
Topic: Charting Output ADC20
Replies: 3
Views: 64

Re: Charting Output ADC20

Please ensure you running the latest version of Picolog 6-
Pico Technology Downloads Page

The 'table view' tab and 'configure data table' button are shown below-
by AndrewA
Wed Jul 31, 2019 1:46 pm
Forum: Getting Started
Topic: Charting Output ADC20
Replies: 3
Views: 64

Re: Charting Output ADC20

Hi, To export to csv click on the 'table view' tab, then click the button 'Configure Data Table' Set the desired settings. You can set the 'Output sample rate' to slower the faster channel to down sample the data. In this case the samples in between can be used to display Average, Minimum and Maximu...
by AndrewA
Mon Jul 29, 2019 4:08 pm
Forum: PicoScope for Linux
Topic: Picoscope 6 install into Lubuntu
Replies: 1
Views: 200

Re: Picoscope 6 install into Lubuntu

Hi,

Can you confirm what version of Lubuntu this is? And if this is a 32-bit or 64-bit version of the OS.
by AndrewA
Fri Jul 26, 2019 8:55 am
Forum: LabVIEW
Topic: PicoScope 4262 streaming in labview
Replies: 2
Views: 140

Re: PicoScope 4262 streaming in labview

Hi Bartek, On the front panel in the streaming example you can set the sample rate under the 'Acquisition Settings' section. Set 'Desired Sample Interval' to 250 and 'Time Units' to ns, giving 250ns = 4MHz sample rate. The actual 'Sample Interval' is returned on the front panel, you may want connect...
by AndrewA
Wed Jul 24, 2019 3:53 pm
Forum: LabVIEW
Topic: Pico 5243D - Sync scope acquisition with AWG signal output
Replies: 5
Views: 1257

Re: Pico 5243D - Sync scope acquisition with AWG signal output

Yes you can do this in the same way with the Rapid Block Bulk example. As before there is no trigger set on the scope. This gives the fastest Rapid Block capture for the scope. I would first adjust scope timebase and number of samples (pre and post) in the 'Rapid Block Acquisition Settings'. Then ad...
by AndrewA
Tue Jul 23, 2019 2:13 pm
Forum: PicoScope 6 for Windows
Topic: Measurement table showing strange values
Replies: 1
Views: 81

Re: Measurement table showing strange values

Hi, The Min, Max, and Average columns are calculated from the last N waveform buffers in the 'capture count' column. You will notice if you stop and restart the scope, the number waveform buffers in top toolbar increase to the maximum set (default is 32) You can change in- 'Tools' menu -> 'Preferenc...
by AndrewA
Thu Jun 13, 2019 4:07 pm
Forum: LabVIEW
Topic: Pico 5243D - Sync scope acquisition with AWG signal output
Replies: 5
Views: 1257

Re: Pico 5243D - Sync scope acquisition with AWG signal output

Here is the working setup and code relating to this post. As this is a question we offer get asked. Hardware setup- Connect the require channels. The AWG output can be optionally connected to a input channel. Software- optionally set the scope trigger settings as per normal, or just leave the trigge...
by AndrewA
Wed Jun 12, 2019 9:39 am
Forum: PicoScope 6 for Windows
Topic: drop-down vert scale issue
Replies: 2
Views: 177

Re: drop-down vert scale issue

Hi,

Does the Dell M4700 support hidpi screen resolution feature? This could be causing the issue.
Here are some possible solutions-
https://windowsreport.com/fix-hidpi-issues-windows-10/

Regards Andrew Abbott
Technical Specialist
by AndrewA
Wed Jun 12, 2019 9:18 am
Forum: General Software Discussions
Topic: ADC20/ADC24 programming guid
Replies: 1
Views: 133

Re: ADC20/ADC24 programming guid

Hi,

An updated version of the programmers guide is now available-
adc-20-24-data-logger-programmers-guide.pdf
by AndrewA
Fri May 03, 2019 12:37 pm
Forum: PicoScope 6 for Windows
Topic: impedance calculation - Power Factor
Replies: 6
Views: 2599

Re: impedance calculation - Power Factor

These maths channels won't in FFT mode in PicoScope 6.
So I would try using the FRA4Picoscope with this new tool-
https://www.eevblog.com/forum/testgear/ ... msg2190516
by AndrewA
Fri May 03, 2019 12:13 pm
Forum: PicoScope for Mac
Topic: Accessing Raw Data
Replies: 2
Views: 488

Re: Accessing Raw Data

Goto the 'Configure Data Table' dialogue by clicking on the table tab, and cog wheel or click 'Configure table' button. Set the table data settings 'Output sample rate' to use the fastest sample rate you have set on your channels. And the then click on 'Last' button to select the latest values for a...
by AndrewA
Fri Apr 26, 2019 10:14 am
Forum: PicoScope 6 for Windows
Topic: impedance calculation - Power Factor
Replies: 6
Views: 2599

Re: impedance calculation - Power Factor

We now have a formula to give a RMS value using a maths channel. sqrt(integral(A^2)/T) Adding '+freq(A)-freq(A)' to it stops the start of the maths channel starting at a lower value. When using the measures on these maths channels you need to set 'between time rulers' and move time rulers to exclude...
by AndrewA
Wed Apr 17, 2019 9:31 am
Forum: USB PC Oscilloscopes
Topic: Detail of how does AWG on Pico 5442B work
Replies: 1
Views: 192

Re: Detail of how does AWG on Pico 5442B work

Note the AWG update rate for PS5000B/D units is 200MHz so this is the maximum frequency that can be generated. The term 'arbitrarywaveform' is your buffer of values/points for the AWG. These are passes as int16_t values, but the AWG resolution for PS5000B/D is 14 bits (output step size approximately...