Oscilloscope Tutorial

Which product is right for your exact requirements
Post Reply
Carlo

Oscilloscope Tutorial

Post by Carlo »

Hi all,

I'm in the market for an oscilloscope. I started reading the Oscilloscope Buyer's Guide: What to look for when choosing an oscilloscope.

In the section labeled as Resolution and Accuracy, how do you get 8 mV per step with an 8 bit resolution?

Also, regarding real time and equivalent time sampling rates, if I have a repetitive signal (e.g. a 1 kHz tone) that is switched off suddenly at intermittent times, do I give priority to the real time sampling rate when picking an oscilloscope?


Thanks!

Michael
Advanced User
Advanced User
Posts: 656
Joined: Thu Jul 07, 2005 12:41 pm
Location: St Neots, Cambridgeshire

Post by Michael »

Hello,

Thank you for your enquiry.

The ±1V range will give you an 8mV step performance with 8bits. The 8bit range will give you a minimum ±100mV range, this would give you 0.78mV steps.

This would largely depend on the length of the full pulse. High memory scopes remove the compromise between capturing the full pulse and reducing the sampling rate. With our new PicoScope 5000 series, the high memory and sampling rate allow you to view slower signals, like your 1kHz tone, over much longer periods of time. This is even possible with the 3000 series. All of this without even switching the ETS mode on!

If you choose a scope with a much higher sampling rate, you can oversample to improve the effective resolution.

Without my advice appearing biased, I would always recommend that you consider future applications as well as current ones when choosing a scope.

If you have any further questions, please don't hesitate to ask.

Best regards,
Michael
Michael - Tech Support
Pico Technology
Web Support Forum

User avatar
Jeff
Advanced User
Advanced User
Posts: 192
Joined: Tue Apr 05, 2005 12:39 pm
Location: St Neots, UK

Re: Oscilloscope Tutorial

Post by Jeff »

Carlo wrote:In the section labeled as Resolution and Accuracy, how do you get 8 mV per step with an 8 bit resolution?
Hi Carlo,

The article you read says that "with a ±1 V range selected", 8 bits gives you 8 mV steps. This was calculated as follows. The voltage swing on a ±1 V range is from -1 V to +1 V, which is a total swing of 2 V. The number of levels at 8-bit resolution is 2^8 = 256. The voltage resolution, or voltage step per level, is therefore 2 V / 256 = 7.8125 mV, or nearly 8 mV.
Jeff