Frequency increment

Post by PhC »

Hello,
I have a problem with the generator of my pico 5444B. The frequency setpoint does not seem to be respected.
The signal from the AWG passes through a 2nd order low pass filter with 0.35Hz cutoff frequency. I do not understand why all frequency deviations are not equal to about 30mHz as expected (see attached document).
pb_Df.pdf
(322 KiB) Downloaded 287 times
thank you in advance.

PhC
User
User
Posts: 2
Joined: Wed Nov 23, 2016 10:58 am

Re: Frequency increment

Post by PhC »

The problem is resolved. The value of increment in frequency was too small compared to the frequency resolution of the generator ... when I set it to 50mHz, all is ok ...

Martyn
Site Admin
Site Admin
Posts: 4491
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: Frequency increment

Post by Martyn »

Good news, and thanks for letting us know.