Error With Driver?

Come here to introduce yourself, who you are, and what you do
Post Reply
dgerrard
Newbie
Posts: 1
Joined: Fri Jul 08, 2016 10:05 pm

Error With Driver?

Post by dgerrard »

My picoscope setup had been working fine but it seems that there is some issue with the driver. When I run this API Function in Matlab I get the following error. I didn't change any code but it quit working (I believe after a Windows update). How can I get this corrected?

[run_block_status, timeIndisposedMs] = invoke(ps3000a_obj, 'ps3000aRunBlock', 0, 0, 4.2950e+09, 1, 0)

run_block_status = 29 (PICO_TOO_MANY_SAMPLES)
timeIndisposedMs = 0
______________________________________________________________________________
Instrument Device Object Using Driver : PS3000a_IC_drv.mdd

Instrument Information
Type: Oscilloscope
Manufacturer: Pico Technology Ltd.
Model: PicoScope 3000 Series

Driver Information
DriverType: MATLAB generic
DriverName: PS3000a_IC_drv.mdd
DriverVersion: 1.0

Communication State
Status: open

Connected to PicoScope device:-
Handle: 16384
Batch/Serial number: CY761/117
InstrumentModel: 3206D
Number of channels: 2
Signal Generator type: None
connect_status =

open_unit: 286
unit_serial: 0
unit_variant: 0

Martyn
Site Admin
Site Admin
Posts: 4491
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: Error With Driver?

Post by Martyn »

The return code 286 from the OpenUnit call means that you have the device connected to a USB2 port (PICO_USB3_0_DEVICE_NON_USB3_0_PORT), can you try on a USB3 port and see if the device is working correctly.