TC-08 uncertainty

Which product is right for your exact requirements
Post Reply
Dan Acres

TC-08 uncertainty

Post by Dan Acres »

Hello, I'm looking at the TC-08 for a high accuracy temperture monitoring system within an enclosure.
I'm not quite clear on what the uncertainty on the TC-08 actually is in high accuracy resolution (0.025K) mode.
According to the online spec the uncertainty in measurement is approx. 1 Kelvin at room temperature. Is this correct?

Sarah

Post by Sarah »

Hi Dan

The accuracy of the TC-08 is the sum of ±0.2% and ±0.5C. The first refers to the measurement and the second to the cold junction measurement. The accuracy of the sensor also needs to be taken into account.

Therefore at room temperature the accuracy would be less than 1C.

Best Regards

Guest

Post by Guest »

Sarah wrote: The first refers to the measurement
In centigrade instead of an absolute measurement?

Sarah

Post by Sarah »

Hi

Yes it is with reference to centigrade.

Best Regards

Dan Acres

Post by Dan Acres »

ok, cheers

Post Reply