3206 sensitivity

Post by rayblane »

I am seriously considering buying a 3206 kit. However, the technical specification states a vertical sensitivity of +/- 100mV to +/- 20V/div. Elsewhere the sensitivity is stated as 10mV to 2V/div. Which is correct?
Also, the stated resolution is 8 bits - are 8 bits used at 100mV and also at 20V?

Michael
Advanced User
Advanced User
Posts: 656
Joined: Thu Jul 07, 2005 12:41 pm
Location: St Neots, Cambridgeshire

Post by Michael »

Hello,

To clarify the resolution, I will show you how to calculate A-D converter steps.

The calculation goes like this with only two variables to change.
A-D mV steps = Voltage range ÷ Bit resolution

Now, you will need to compute the bit resolution like this: 2x10^n where n=bit resolution. Next you will need to compute the voltage range. A ±500mV range covers -500mV to +500mV, a range of 1000mV.

So the calculation now looks like this:
3.9mV = 1000mV ÷ 256 (2x10^8)

Hope this makes sense.

Best regards,
Michael
Michael - Tech Support
Pico Technology
Web Support Forum

User avatar
Carl Grotti
Active User
Active User
Posts: 19
Joined: Thu Nov 21, 2002 5:17 am
Location: Missouri, USA

Post by Carl Grotti »

Hello Michael,

I hope you don't mind me jumping in here.

Your step calculation makes perfect sense except for one thing. How does 2x10^8 =256? I'm not familiar with the "^" symbol.

Thanks,

Michael
Advanced User
Advanced User
Posts: 656
Joined: Thu Jul 07, 2005 12:41 pm
Location: St Neots, Cambridgeshire

Post by Michael »

Hello Carl,

Sorry for my lack of ASCII codes, I was trying to represent 2 power 8 in symbol and used the carat symbol for lack of superscripts.

The calculator version would be like this: (2^8)
3.9mV = 1000mV ÷ 256 (2^8)

Sorry for the error.

Regards,
Michael
Michael - Tech Support
Pico Technology
Web Support Forum

rayblane

3206 resolution

Post by rayblane »

Thanks for the replies. You have answered the question about resolution, but not about sensitivity.
Which of the two quoted vertical deflection ranges is correct, 10mV to 2V or 100mV to 20V?

Thanks.

Sarah

Post by Sarah »

Hi

Thank you for your post.

The correct ranges are from 100mV to 20V - the maximum input is ±20V.

Best Regards