Post by Asim »

Q-1
I want to measure frequency & amplitude of a pulse having width of 13msec and time interval between two pulses is 1.7sec. The frequency of the pulse is 9.5kHz.

Q-2
Can we use ADC-212 to record and analyze the pulse train?

Thanks
Dr. Asim

User avatar
matthew
Advanced User
Advanced User
Posts: 109
Joined: Wed Sep 25, 2002 9:35 am
Location: Cambridgeshire, UK

Repetitive Pulse Trains

Post by matthew »

Thank you for the post Dr Asim,
I have moved this topic to the Pre-Sales Advice section since this is more appropriate.

The ADC-212/100 will meet your requirements:
The Analogue bandwidth of the ADC-212/100 is 50Mhz (you only need 10kHz).
In order to achieve the 1.7 second interval time we will need to work out the memory buffer's download time. The memory buffer can download at a rate of 60,000 Samples per second. Therefore we can take no more than 1.5 x 60,000 = 90,000 Samples over the 13 ms period (equates to approx. 7M Samples per second).

In short, this will be no problem for the ADC-212/100.

Best Regards,
Matt Everett

Pico Software Engineer

mblaow

ADC-216

Post by mblaow »

Can you please tell me the price of ADC-216 ?

User avatar
matthew
Advanced User
Advanced User
Posts: 109
Joined: Wed Sep 25, 2002 9:35 am
Location: Cambridgeshire, UK

Post by matthew »

Hi,

The current price of the ADC-216 is £499GBP excluding tax and delivery.

Best Regards,
Matt Everett

Pico Software Engineer