Spectrum analyser update rate at lower spans

Forum for discussing PicoScope (version 5)
Post Reply
Guest

Spectrum analyser update rate at lower spans

Post by Guest »

On my 3224, why does the spectrum analyser update rate (sweep time) reduce dramatically at lower spans? When I look at the bottom few hundred hz of the spectrum, the update rate is of the order of a few seconds. Intuitively, I would expect the update rate to be faster for lower freqs and higher for higher freqs.

V5.13.7

Jonathan

Sarah

Post by Sarah »

Hi

Thank you for your email.

These lower frequencies take longer to capture fully as they by definition have a longer wavelength. Therefore the scan time needs to be longer and thus the update takes a bit longer. This is normal when using the spectrum analyser.