Call to ps3000_ready does not return

Post general discussions on using our drivers to write your own software here
Post Reply
cstook
Newbie
Posts: 0
Joined: Mon Dec 21, 2015 4:42 am

Call to ps3000_ready does not return

Post by cstook »

I an using the beta version of 64 bit windows SDK (PicoBetaSDK_64_10.6.11.23). Operating system is Windows 10. Scope is Picoscope 3425.

ps3000_get_unit_info returns:
PS3000_DRIVER_VERSION = 3.7.6.4
PS3000_USB_VERSION = 2.0
PS3000_HARDWARE_VERSION = 8
PS3000_VARIANT_INFO = 3425
PS3000_BATCH_AND_SERIAL = XFY20/041
PS3000_CAL_DATE = 16Jul09
PS3000_ERROR_CODE = 0

I am using the scope in block mode. Sometimes calls to ps3000_ready do not return, usually after several thousand successful calls. Opening the chrome browser increases the probability of ps3000_ready not returning (while chrome is opening, not after it is open). As a temporary solution, I replaced polling ps3000_ready with a delay > time_indisposed_ms before calling ps3000_get_values. This works OK, but I prefer to not have my code dependent on delays.

Using a delay before calling ps3000_ready, so that ps3000_ready always returns true, does not fix the problem.

This looks similar to the problem seen by Edk in 2013 (post35621.html#p35621).

Is there a solution?

AndrewA
PICO STAFF
PICO STAFF
Posts: 394
Joined: Tue Oct 21, 2014 3:07 pm

Re: Call to ps3000_ready does not return

Post by AndrewA »

Hello cstook,

I have tested this with C console example the yes the issue seems to be with the 3425 and 3424 and possibly the 3224.
I also tested a 3204 this did not seen to have the issue.
Yes- adding a delay (Sleep(100);) before ps3000_stop() or ps3000_get_values() seems to fix the issue.
I have created a ticket for this, and will await a response from our development team.
Regards Andrew
Technical Specialist