5244B - Channel A and Channel B not equal

Post any questions you may have about our current range of oscilloscopes
Post Reply
Moonwalker
Newbie
Posts: 0
Joined: Tue Feb 11, 2014 6:36 pm

5244B - Channel A and Channel B not equal

Post by Moonwalker »

I've been measuring a signal from a buck converter and noticed that channel A and channel B are not showing exactly the same signal. Channel B is showing a larger ripple than channel A. See file attached.
Any help?
Attachments
20151017-0002.psdata
(434.29 KiB) Downloaded 508 times

Martyn
Site Admin
Site Admin
Posts: 4491
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: 5244B - Channel A and Channel B not equal

Post by Martyn »

If you swap the probes around do the differences stay with the channel or follow the probe ?
Martyn
Technical Support Manager

Moonwalker
Newbie
Posts: 0
Joined: Tue Feb 11, 2014 6:36 pm

Re: 5244B - Channel A and Channel B not equal

Post by Moonwalker »

Ohh... sometimes the mind forgets the simple things :)

It looks like it was the probe picking up more signal inductively. After carefully placing the probes close together I get same signal on both channels.

So is it normal for a probe to pick up some signal inductively when connected in circuit?

AndrewA
PICO STAFF
PICO STAFF
Posts: 400
Joined: Tue Oct 21, 2014 3:07 pm

Re: 5244B - Channel A and Channel B not equal

Post by AndrewA »

Yes the scope probes can pickup RF if the signal is a high signal level.
The probe must have been picking up the RF from the buck converter itself.
Regards Andrew
Technical Specialist