Signal generator switching frequency

Post by Inverterguru »

Hello,

I want to switch the signal generator frequency. To test it I create the attached VI. If I run the VI there are drops in the signal wave (see attached oscillogram). How I can switch the frequency without these drops?
(I don't want to build a sweeping generator. I create the VI as a sweeping mode only for testing).

Best regards,
Ulrich

LabView 2014
PicoScope 3406D MSO
PicoScope3000a.dll (32bit)
Attachments
VI as Picture
VI as Picture
Oscillogram
Oscillogram
SCR0001.png (2.4 KiB) Viewed 5794 times
Bode.vi
Original VI (LabView 2014)
(12.48 KiB) Downloaded 528 times

AndrewA
PICO STAFF
PICO STAFF
Posts: 400
Joined: Tue Oct 21, 2014 3:07 pm

Re: Signal generator switching frequency

Post by AndrewA »

Hello Ulrich,

I would remove the 100ms delay. You may still find you have a vertical edge in the waveform when changing frequencies.
Regards Andrew
Technical Specialist

Inverterguru
Newbie
Posts: 0
Joined: Thu Oct 15, 2015 10:13 am

Re: Signal generator switching frequency

Post by Inverterguru »

Hello Andrew,

if I remove the delay the drops come up more. I find out that these is a problem of the firmware, not a problem of LabView or an other language. If I change the frequency with the PicoScope software the same problem occurs.

Thank you for your help.

Best regards,
Ulrich

AndrewA
PICO STAFF
PICO STAFF
Posts: 400
Joined: Tue Oct 21, 2014 3:07 pm

Re: Signal generator switching frequency

Post by AndrewA »

Yes this not avoidable, its just how the hardware of the signal generator works. You could setup a sweep with the signal generator if you want to step through a range of frequencies.
Regards Andrew