how to set sample clock timebase rate

Post any questions you may have about our current range of oscilloscopes
Post Reply
freemason
Newbie
Posts: 0
Joined: Fri Aug 14, 2015 10:54 pm

how to set sample clock timebase rate

Post by freemason »

Hello,
I am using Picoscope 4824, and have a requirement of setting a sample clock of 6.67MHz, how would i set up this?

i have the following setting on national instruments card and i want to replicate this on picoscope
Device: DAQ::1 (PCMCIA-5102)
Sample Rate (S/s): 6.667M
Record Length: 1.051k
Acquisition Mode: N Samples
Configure acquisition settings.png
Thanks

Martyn
Site Admin
Site Admin
Posts: 4499
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: how to set sample clock timebase rate

Post by Martyn »

The master clock of the 4824 is 80Mhz, and the possible sampling frequencies are related to this value, defined by the equation 80/(n+1) where n is the timebase. So for a timebase of 11 the sampling frequency will be 80/(11+1) or 6.6666666... MHz

So provided this is the frequency you actually require, and not specifcally 6.67MHz, then all you would need to do is select a timebase of 11 in the RunBlock function call.
Martyn
Technical Support Manager