User of Picoscope4223-14 problems with Duty cycle messuremen

Having problems ? let us know the details here
Post Reply
Adriaan
Newbie
Posts: 0
Joined: Wed Jan 22, 2014 9:11 am

User of Picoscope4223-14 problems with Duty cycle messuremen

Post by Adriaan »

This is the measurement that I have made.
This is the measurement that I have made.
Hello,

I have problems taking measurements with the function Duty Cycle with our new PicoScope. The problem is that it is a signal taken of a coil wich produce an induction. It is a automotive signal of the EVAP. I think it take the Duty Cycle time of the induction peak and not the time of the actuation and deactuation of the signal. Can you help me with this?

Martyn
Site Admin
Site Admin
Posts: 4499
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: User of Picoscope4223-14 problems with Duty cycle messur

Post by Martyn »

Try Tools->MathsChannels and enable Invert A, and then perform the duty cycle calculation on this, it will be inverted so needs subtracting from 100 to give the value you require.

A bit of a work around, but I will submit a request to the team to add a threshold level to perform duty cycle calculations on.
Martyn
Technical Support Manager

Adriaan
Newbie
Posts: 0
Joined: Wed Jan 22, 2014 9:11 am

Re: User of Picoscope4223-14 problems with Duty cycle messur

Post by Adriaan »

Hello Martin,

I will try this later on. thanks for your supply.

have a good weekend,

Adriaan