National Standards Calibration Service

Which product is right for your exact requirements
Post Reply
Gustav
Newbie
Posts: 1
Joined: Sat Jan 23, 2010 1:33 pm

National Standards Calibration Service

Post by Gustav »

Hi,

I am thinking very much about buying a 5244B. I am not sure if I need the calibration service. The only reason I can think of is the timebase accuracy (drift). The datasheet states that it is ±2 ppm (±1 ppm/year).

I want to know, how much more precise does it get? Is the NS calibration service able to measure the timebase below 1ppm? Maybe to +/-0.1ppm?

Martyn
Site Admin
Site Admin
Posts: 4491
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: National Standards Calibration Service

Post by Martyn »

The calibration will tell you the actual error at the time the calibration is carried out, you can then account for the errors in any measurements you make.
All our scopes are, of course, calibrated during final test and are guaranteed to meet the published specification for but no calibration certificate is made or issued. If a calibration certificate is requested from new, or a scope is later returned for calibration, the actual errors & accuracy for that particular scope are measured and recorded on the certificate. For example, a scope might have a published accuracy of +/-1% for voltage. A scope direct from the factory will meet or better that accuracy. A calibrated scope will have the actual error recorded for each voltage range and that error might be a reading that is 0.5% high. This error can now be allowed for when taking accurate measurements.
Martyn
Technical Support Manager