Probe attenuation setting

Post your LabVIEW discussions here
Post Reply
ulek
Newbie
Posts: 0
Joined: Tue May 28, 2013 8:52 am

Probe attenuation setting

Post by ulek »

Hey all,

How would I go about setting custom attenuation for the probes on the picoscope in labview?

When I use 100 probes then the signal in the labview is absolutely ugly.

Thanks
Ulek

Hitesh

Re: Probe attenuation setting

Post by Hitesh »

Hi Ulek,

You will need to scale your readings back e.g. multiply by 100 if you are using the x100 setting.

I hope this helps.

Regards,

ulek
Newbie
Posts: 0
Joined: Tue May 28, 2013 8:52 am

Re: Probe attenuation setting

Post by ulek »

Hey,

That is what I am doing right now.

When I use the 10x probes and account for that (after digital to analog conversion multiply by 10) in the program everything is fine.

When I use the 100x probes and do the same thing as for the 10x probes (multiply 100 instead of 10) the signal is extremely noisy and basically unusable.

Hitesh

Re: Probe attenuation setting

Post by Hitesh »

Hi Ulek,

What is the amplitude of the signal you are measuring? Which PicoScope and 100x probe are you using?

It might be that the attenuation is taking the signal close to the noise floor. Have you tried changing the voltage range?

Regards,