Pressure monitoring

Which product is right for your exact requirements
Post Reply
Dr D Patel
Newbie
Posts: 1
Joined: Tue Nov 02, 2004 4:16 pm

Pressure monitoring

Post by Dr D Patel »

I am willing to capture the pressure data from a transducer/transmitter for continuous monitoring. the out put signals are in the range of 4-20mA. We are also planning to get the captured information being transmitted to SCADA system.

Please advise on the right product for us.

User avatar
markspencer
Site Admin
Site Admin
Posts: 598
Joined: Wed May 07, 2003 9:45 am

Post by markspencer »

Hi,

Thank you for your post.

It is possible to measure 4-20mA with our equipment. This can be achieved by using a shunt resistor between gnd and channel. The software can the use scaling files to convert it to a pressure unit. I would recommend that you look at the ADC-11/12, this is an 11 channel unit, 12 bits of resolution. There is also a terminal board available with screw terminals to make connecting your cables easier. The shunt resistor can also be placed onto the terminal board.

Unfortunately, I am not familiar with SCADA systems, but our software does not link with other systems. The ADC-11 has to be connected to a PC at all times and the software needs to be running at all times whilst in use.

It is possible to use the drivers to write your own software if you wished to go down this route.

Best regards,
Regards,

Mark Spencer