Picolog1216

Post any questions you may have about our current range of USB data loggers
Post Reply
picolog1216
Newbie
Posts: 0
Joined: Thu Jul 19, 2012 12:52 pm

Picolog1216

Post by picolog1216 »

I purchased and am currently using Picolog 1216. I need to measure 16 voltages simultaneously. The voltage range is around 2.2 to 2.5 mV. When I look at the terminal board, i see only six ground sockets. How can I measure sixteen voltages simultaneously? Any help? Thanks.

Martyn
Site Admin
Site Admin
Posts: 4491
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: Picolog1216

Post by Martyn »

Are the ground reference points for the sixteen voltages the same ?
Martyn
Technical Support Manager

picolog1216
Newbie
Posts: 0
Joined: Thu Jul 19, 2012 12:52 pm

Re: Picolog1216

Post by picolog1216 »

no, they are not at the same ground state.

Martyn
Site Admin
Site Admin
Posts: 4491
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: Picolog1216

Post by Martyn »

If you are not able to common the grounds then the 1216 would not be correct for this situation. If you are able to then attach multiple grounds to each ground socket.
Martyn