cepe

Alarm holdoff, which units to use?

Post by cepe »

Thanks for fixing the power calculation problem in 5.10.1 quickly.

I am now fiddling with setting the alarm holdoff. Which unit is the holdoff mesaured in? Micro/milliseconds or # of samples of the recording samling rate.

BTW, I have read the documentation without finding an answer. I am using a DrDaq.

Regards,

--cepe
Oslo, Norway
Outside temp: 5,2C
Outside RH: 86,7%

cepe

Another question, handling out of range situation with DD101

Post by cepe »

I measure humidity with DD101. When the humidity mesured exceeds the 87-88% range, values are reported as "*****", which I assume is indicating "sensor failed". I would like the values to just be the last valid measurement, or clipped to max/min range value.

Can I/how can I do that?

--cepe

User avatar
markspencer
Site Admin
Site Admin
Posts: 598
Joined: Wed May 07, 2003 9:45 am

Post by markspencer »

Hi,

Thank you for your request, the HoldOff is in seconds.

Best regards,
Regards,

Mark Spencer