I have some problems getting the right measurements from ADC11 to C++.
I measure temperature with af termistor within the rage of 0 - 700 mV
When I see the recording in the picorecorder the mV increases steady in a secure and fine way, but when I see the mV in my C++ programme its start fine when the rise in temperature is high, but when the rise i slowly the mV i jumping up and down (f.ex mV: 522, 527, 518, 529,535, 528 .....). When the picorecorder is measuring at the same time, it wil steady increase, whitout jumping up and down (f.ex. V: 0.522, 0.527, 0.529, 0.529, 0.535, 0.538 ....).
I looks like I am not getting the right signal ???
The driver that you are using for your C++ program and the driver that is being used with Picolog ar one and the same. Check that you are using the same version, this can be done via the date time properties.
When you are collecting the readings with you own application are you collecting it at the same time interval in Picolog. Check whether you have it set to 'single' or 'take as many samples as possible', this can be found in the 'Sampling Rate' window.