DrDAQ indicate wrong pH

Forum for discussing PicoLog
Post Reply
kiyoshi_abe
Newbie
Posts: 1
Joined: Thu Sep 13, 2007 9:43 am

DrDAQ indicate wrong pH

Post by kiyoshi_abe »

Hello,
I use DrDAQ for pH measurements and the readings are not correct with the picolog recorder, where the reading is stuck in a value around 7 in all cases. Moreover, readings shows 7.56 in 0.1mol/L HCl solution, and 6.74 is shown in 0.1mol/L NaOH solution.

This situation is quite the same also in Parallel Port and USB Parallel Port
Adapter.

I use picolog recorder rerease R5.20.1 and Windows Vista Home Premium SP1.

Could you please offer a solution?

Thank you

picojohn
Advanced User
Advanced User
Posts: 382
Joined: Mon Jun 11, 2007 1:10 pm

Post by picojohn »

Hello kiyoshi_abe,

It is possible that your pH probe is defective as they can dry out over time and, the easiest check would be to use a known working probe in a buffer solution of known pH.

Regards
John

Guest

DrDAQ indicate wrong pH

Post by Guest »

Thank you.
Each solutions of 0.1mol/L HCl(Factor 0.999) and 0.1mol/L NaOH(Factor 1.002) are both volumetric analysis solution. I am using these solutions for the neutralizing titration.When I measured them by other pH meter, they indicate 1.01 and 13.01 respectively.
DrDAQ displays a wrong pH though the pH probe of this pH meter is substituted for DrDAQ's one.
It seems that this pH probe is not out of order.

When I use picolog recorder release R5.08.6 and Windows Vista Home Premium SP1, DrDAQ indicate 1.10 in 0.1mol/L HCl solution and 12.30 in 0.1mol/L NaOH solution.
However, picopp.sys and DRDAQ32.DLL were substituted as follows.

picopp.sys 2002/07/31 => 2007/07/02 (R5.20.1)
DRDAQ32.DLL 2002/07/31 => 2008/05/02 (R5.20.1)

Best Regards.

ziko
Advanced User
Advanced User
Posts: 1705
Joined: Fri Dec 01, 2006 10:03 am
Location: St Neots

Post by ziko »

Hi sorry for the delayed response, John does not work with us anymore.

There may be a problem your DrDaq. Please email us at support@picotech.com so we can look into this further.

Kind regards.
Ziko

Technical Specialist

ziko
Advanced User
Advanced User
Posts: 1705
Joined: Fri Dec 01, 2006 10:03 am
Location: St Neots

Post by ziko »

Actually one other thing you can try before sending us an email is to short out the BNC connector using a terminator plug (or paper clip). The reading on the screen should be pH 7 (0.5).

Kind regards
Ziko

Technical Specialist