Buffer data aquisition timing

Post by pyjamalady »

Hello,

I just started to use a 2204A to characterise fields stability and so on in ultrasonic cleaning baths.
When I stored a data with the original settings I got a set of 32 buffers data.

I would like to know these buffer data aquisition timing and/or data aqision timing criteria for each data.

I tried to seach these information but could not find.

Any help is much appreciated
Pyjamalady

Martyn
Site Admin
Site Admin
Posts: 4491
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots

Re: Buffer data aquisition timing

Post by Martyn »

The 32 buffers is the default setting for the maximum waveforms (buffers) in Tools->Preferences

This is the number of buffers you would like to capture in a run before the oldest buffer is overwritten for each new collection. The actual number of buffers you can collect will depend upon the setup, specifically the time across the screen, the number of samples requested per buffer, and the number of active channels, and may be less than this value.
Martyn
Technical Support Manager