I am trying to use adc200.vi (parallel port) in LabView 8.0 on a Windows XP system. Whenever I try to run it comes back with the error message -VI "adc200.vi" was stopped at Call Library Function Node 0x2074 of subVI "adc200.vi"". Can you help me please.
I have had a look at this VI and found a couple of problems.
Fisrtly, the ADC200_run function was causing a crash as there was a missing returns declaration. Next, there was no ADC200_close function call which was preserving the driver within the LabVIEW development enviroment. The only way to escape this was to restart LabVIEW. I have now added a close function at the end of the stacked sequence.
This should help you out, you will need to register and post a reply to see the attachment in the topic review window.
Go to the post reply section and scroll down. You will see the topic review section and then scroll down to through the messages until you see my message with the attachment.
Best regards,
Michael
Michael - Tech Support
Pico Technology
Web Support Forum
I'm having the same problem with my .vi, however, I have Labview 7.1, and the version posted here is for 8.0. Would it be possible to get the updated .vi in a version compatible with 7.1?