High speed Temp logging using TC08(USB) via LabVieW 8.2

Post by vavoul »

Dear All,

I am trying to write a vi for a TC08(USB) that will enable me to acquire K thermocouple data at ~high speed over 30 Hz per channel. The Usage of labview is mandatory since I have to integrate (synchronized) this vi with other DAQ and DMM measurements.

From manual I understand that in order to succeed high data logging rates the streaming mode should be selected via the use of usb_tc08_run and usb_tc08_get_temp commands. Unfortunately, all my efforts are unsuccesfull and I desparately want help. :)

Using Get Single mode, I managed to prepare a vi tha suits my application logic perfectly but the acquisition rate is too slow only 4 Hz (even for two channels). ( I attach the vi: USB_TC-08_Single_82_2channels_4Hertz.vi)

Additionally I attach the outcome of my efforts for streaming mode. I tried to keep same logic since this suits my experimental plan, but when I start the vi, LABVIEW closes....with the standard message :"LabVIEW 8.2 Development System has encountered a problem and needs to close. We are sorry for the inconvenience."

Thank you in advance,

Antonios
Attachments
USB_TC-08_Single_82_2channels_4Hertz.vi
succesfull effort for singe get mode
(90.76 KiB) Downloaded 758 times
USB_TC-08_Single_82_2channels_streaming.vi
effort for streaming mode
(22.96 KiB) Downloaded 744 times

gruntman
Advanced User
Advanced User
Posts: 109
Joined: Thu Sep 28, 2006 3:50 pm

Post by gruntman »

Antonios,

The TC-08 has a conversion time of ~100ms per channel. Therefore even if only one channel was enabled you will be able to only collect data at approximately 10Hz under ideal conditions, nowhere near your desired 30Hz.

I briefly looked at your streaming example and noticed one major flaw. You are calling usb_tc08_legacy_get_temp when you should be calling usb_tc08_get_temp. The legacy method only works if you set the TC-08 up into legacy mode which you do not need to do.

Regards,

Richard Boyd
Crag Technologies, Inc
http://www.pc-oscilloscopes.com