Test and Measurement Forum

usb TC-08 reading synchronization

Forum for discussing PicoLog

usb TC-08 reading synchronization

Postby fc_ » Tue Dec 12, 2017 10:55 pm

hello, i'm using the usb tc-08 and Picolog in an application with relatively fast varying temperatures (welding).
I'm aware of the fact that the cycle time, with N thermocouples, is 100*(N+1) milliseconds.
What is still unclear, for me, is in which instant each individual thermocouple is read.
A more clear knowledge of the functioning of the unit could permit to interpolate the values and overcome the problem of the relatively slow complessive sampling rate.
Thanks
fc_
Newbie
 
Posts: 0
Joined: Sat Oct 26, 2013 10:50 am

Re: usb TC-08 reading synchronization

Postby Martyn » Mon Dec 18, 2017 8:58 am

It is sequential, once you start sampling the first sample will be after 100ms, which will be for the first enabled channel, then after the next 100ms, the second enabled channel, ..... and finally the cold junction before the process repeats.
Martyn
Technical Support Manager
Martyn
Site Admin
Site Admin
 
Posts: 2851
Joined: Fri Jun 10, 2011 8:15 am
Location: St. Neots


Return to PicoLog

Who is online