40 Channels - 10 Hz - 15 min - TC Typ K

Post by Guest »

Hello,

which system do I need to perform the acquisition descriped above?
Channels: 40
Frequency: 10 Hz
Duration: 15 minutes
Input: Thermocouple Typ K

Thanks in advance!

Guest

Changed sample rate

Post by Guest »

Which system will fullfill the requirments best, when changed the sample rate to 1 Hz?
Thanks!!

ziko
Advanced User
Advanced User
Posts: 1705
Joined: Fri Dec 01, 2006 10:03 am
Location: St Neots

Post by ziko »

Hi and thank you for your post.

You cannot achieve the 10Hz sample rate, however you can achieve the 1Hz sample rate. We have the USB-TC08, this is an 8 channel thermocouple logger, you can have multiple loggers connected to one PC logging using one instance of the software.

http://www.picotech.com/thermocouple.html

***Hi just to add you will not get 1Hz sampling updates with multiple devices. You can if you have one TC-08, however with multiple devices there are other associated overheads, it is difficult to give an exact figure***

Hope this helps.

Kind regards
Ziko

Technical Specialist