PicoScope 7 Software
Available on Windows, Mac and Linux
Andreas,Andreas_ wrote:Hello Aaron,
im using the 5444A.
My 2nd question was not to turn off the whole numbering of the axis.
I do simply not want 2 same numbers on different places on the y-axis.
So I need either a further digit (-41.15) or leaving off every 2nd number of the y-axis.
Otherwise the Y-axis numbering is somewhat confusing.
(see attachment).
with best regards
Andreas
Thank you Martyn. I'll take a look. If I understand you correctly, to handle the general case, the base/reference number carries some number of the significant digits. So, techniques like that (including use of scientific mode) can help, but would still have a limit. So I guess it's a matter of how much more precision to be able to show.Martyn wrote:In the example you could reference against -40 and then just display the offset from this.
Take a look at the way the PicoScope 6 software handles zooming in so that you always display 4 relevant digits.
Hi Bernhard,mrtl wrote:Hi Aaron,
we would like to use the PicoScopes instead of Velleman equipment for our practical student curses in the future. Because of this, the FRA (Bode Plot) is very important for us.
Unfortunately i get the same error than Martin.S, when using a PicoScope 3203D:
Status: 3203D S/N: ES292/096 successfully initialized.
Status: Starting frequency step 1 (9.998 Hz)
...
Status: Starting frequency step 31 (10000.002 Hz)
Fatal error: Failed to setup stimulus signal: 46
Error: Failed to disable signal generator.
I already installed the 32-bit SDK, but without success.
Kind regards
Bernhard
Thank you for the update, Andreas. It's very helpful to know which scopes are compatible.Andreas_ wrote:Hello Aaron,
don´t know wether you have a PS4824 in your compatibility list.
Today I could do a quick check at the device from work.
I also have the error code 46 on this device.
But the bode plot looks to be ok.
Compared to the PS5444A-plot some days ago.
With best regards
Andreas
Hi Aaron,hexamer wrote:Hi Bernhard,mrtl wrote:Hi Aaron,
we would like to use the PicoScopes instead of Velleman equipment for our practical student curses in the future. Because of this, the FRA (Bode Plot) is very important for us.
Unfortunately i get the same error than Martin.S, when using a PicoScope 3203D:
Status: 3203D S/N: ES292/096 successfully initialized.
Status: Starting frequency step 1 (9.998 Hz)
...
Status: Starting frequency step 31 (10000.002 Hz)
Fatal error: Failed to setup stimulus signal: 46
Error: Failed to disable signal generator.
I already installed the 32-bit SDK, but without success.
Kind regards
Bernhard
Can you confirm whether you are able to produce a Bode plot that looks as expected?
The error messages you show (despite their ominous wording) are harmless. Since this topic tends to come up fairly often, let me try to explain some more details:
A while back it was reported that after a FRA run the signal generator remains active at the last tested frequency. I attempted to resolve this problem, but there is no unified SDK API to disable the signal generator. So I tried to do the next best thing and set the signal generator to 0V, 0 Hz. I set it to 0Hz, because on the PS3000 SDK that should disable the signal generator. Unfortunately that input combination on other scope family SDKs results in an error returned by the API, which the FRA application reports. In general (throughout the actual FRA execution), failing to set the signal generator should be a "fatal error" because the FRA would not be correct. In the case of attempting to disable the signal generator, it should not be fatal because the result is just that the signal generator remain on.
I had been waiting on an API enhancement to fix all of this, but it's not available yet. topic14569.html
Thank you,
Aaron.
Bernhard,mrtl wrote:Hi Aaron,
i made a small electric circuit with a resistor an a condenser in series connection.
With the Velleman i get a correct diagram, but not with the PicoScope.
Just look at the below pictures. Kind regards Bernhard
But my response plot doesn't looks so smooth like in the last post aboveFatal error: Failed to setup stimulus signal: 46
Error: Failed to disable signal generator.
But my response plot doesn't looks so smooth like in the last post aboveFatal error: Failed to setup stimulus signal: 46
Error: Failed to disable signal generator.
Hello and thanks for reporting your issue.meiernt wrote:Hi,
we use an PS3206B. I also get this warning at the end:
But my response plot doesn't looks so smooth like in the last post aboveFatal error: Failed to setup stimulus signal: 46
Error: Failed to disable signal generator.
I connected an simple RC (1 kOhm, 1uF) as Low-pass filter between output and input A.