hexamer wrote:martin.s wrote:Hello,
I have a problem. FRA4 not works with my new 2208B - "Error: Unable to initialize device with serial number EY465/0001"
I am using PicoSDK_64_10.6.10.24 from included CD.
Before there I used 2204A and everything worked perfect.
Can you tell me some reason? I tried to use 32bit sdk or old version 64bit and something else and nothing works.
I need to do a bode plot with FRA4.
Can you help me? Thanks.
Hello and thank you for reporting this - I suspect this may be a common problem. Since the SDK does not yet include support for the newer 2000B scopes, the API DLLs have to come from the PicoScope application. The FRA4PicoScope application prefers to get its DLLs from the SDK location. So it will load incompatible DLLs for these newer scopes. The workaround is to uninstall all PicoScope 32 bit SDKs, keep the PicoScope Application installed, then run the FRA application. You should be able to keep 64 bit SDKs installed if you need them for something else.
Thanks You,
Aaron.
martin.s wrote:hexamer wrote:martin.s wrote:Hello,
I have a problem. FRA4 not works with my new 2208B - "Error: Unable to initialize device with serial number EY465/0001"
I am using PicoSDK_64_10.6.10.24 from included CD.
Before there I used 2204A and everything worked perfect.
Can you tell me some reason? I tried to use 32bit sdk or old version 64bit and something else and nothing works.
I need to do a bode plot with FRA4.
Can you help me? Thanks.
Hello and thank you for reporting this - I suspect this may be a common problem. Since the SDK does not yet include support for the newer 2000B scopes, the API DLLs have to come from the PicoScope application. The FRA4PicoScope application prefers to get its DLLs from the SDK location. So it will load incompatible DLLs for these newer scopes. The workaround is to uninstall all PicoScope 32 bit SDKs, keep the PicoScope Application installed, then run the FRA application. You should be able to keep 64 bit SDKs installed if you need them for something else.
Thanks You,
Aaron.
Hello Aaron,
I tried to use a new PicoSDK and PicoScope app today morning.
I uninstalled everything before. FRA4 still have a problem with my 2006B: Error: Unable to initialize device with serial number EY465/0001.
Can you help me with that?
Yours sincerely
Martin.
hexamer wrote:Hi Martin,
I'm afraid I'm having difficulty replicating this problem. When I install the new 32 bit SDK (version 10.6.12.41), without even re-linking the application (using v0.6b as supplied), I can use my PS2408B just fine. The only difference between my scope and your is two vs four channels. As far as I can tell the application should be able to use a PS2208B scope.
Could you please confirm the following for me:
1) You installed the 32 bit version of the new SDK
2) Your scope still works with the PicoScope Application.
Aaron.
ingmarp wrote:Hello,
I'm new on this blog and have almost the same issue on my PS 2200A: although I can clearly see it in the Select Device box with the correct serial number, the message i get is unfortunately the same "cannot initialize the device".
Which is a pity as.
I'm running FRA on Win7 and tried both FRA 0.6 and debug version.
Thanks and regards.
Ingmar
hexamer wrote:Martin,
I've released a version (0.6.1b) which I hope will resolve the issue you reported. Please install and let me know if it works:
https://bitbucket.org/hexamer/fra4picos ... 00.6.1.msi
Ingmar,
This version might solve the issue you reported too. It looks like I somehow also missed all scopes in the PS2200A family.
Thank you for taking the time to report these issues and for your patience while worked on fixing them.
Aaron.