Pt100 2-wire not selctable

Having problems ? let us know the details here
Post Reply
laser-axel
Newbie
Posts: 1
Joined: Fri Jun 02, 2017 7:27 am

Pt100 2-wire not selctable

Post by laser-axel »

Hello,

I am trying to use a PT-104. The recorder software is 5.25.3. Operating System is windows 10.
I have a pt100-sensor in 2-wire wiring.
In the software I can only select 3- or 4-wire???
The option for 2-wire is simply not present.

Gerry
PICO STAFF
PICO STAFF
Posts: 1145
Joined: Mon Aug 11, 2014 11:14 am

Re: Pt100 2-wire not selctable

Post by Gerry »

Hi laser-axel,

The latest version of PicoLog in English does have the 2-wire option, so there may be an issue, for instance, in the translation to alternative language versions. In the meantime, you could just use the 4-wire version, because you will be making the same connections anyway. (If you must use them, you can make the 2-wire pin connections, as suggested on Page 9 of the PT-104 Data Logger User guide here: https://www.picotech.com/download/manua ... 4.en-2.pdf.)

We don't generally recommend using 2-wire RTD sensors with the PT-104 because the length of the connector cable will affect the accuracy of the temperature readings that you will get.

Regards,

Gerry