Pressure Transducer for 2202

Post by Guest »

I wish to use one of the channels to log out from a pressure transducer.

If I use a 4-20 mA output type, what resistance should I use to give me the preferred input volt range to the data logger?

Alternatively what is the best voltage output for a transducer with a voltage out put?

ziko
Advanced User
Advanced User
Posts: 1705
Joined: Fri Dec 01, 2006 10:03 am
Location: St Neots

Post by ziko »

Hi and thank you for your post.

The 2202 has the following voltage ranges:

+/-50mV
+/- 100mV
+/- 200mV
+/- 500mV
+/- 1V
+/- 2V
+/- 5V
+/- 10V
+/- 20V

It is 8bits in each of these modes.

Hope this helps.

Kind regards
Ziko

Technical Specialist