PicoScope 7 Software
Available on Windows, Mac and Linux
PicoLog CM3
The PicoLog CM3 Current Data Logger is a compact, easy-to-use data logger for measuring the current consumption of buildings and machinery.
PicoLog CM3 current data logger specifications | |
---|---|
Number of channels | 3 |
Range (voltage input) | 0 to 1 V AC RMS |
Accuracy (voltage input) < 200 mV RMS < 1 V RMS |
±1% ±2.5% |
RMS noise | 60 µV |
Resolution | 24 bit |
Conversion time per enabled channel | 720 ms |
Voltage input | 1 V RMS, 20 Hz to 1 kHz |
Input connectors | 4 mm sockets |
Input impedance | >1 MΩ, AC coupled |
Overvoltage protection | ±30 V DC |
Environmental | |
---|---|
Operating environment Temperature range Humidity range |
0 °C to 50 °C (20 °C to 28 °C for stated accuracy) 20% to 80% RH, non-condensing |
Storage environment Temperature range Humidity range |
–20 °C to +80 °C 5% to 95% RH, non-condensing |
Physical properties | |
---|---|
Dimensions | 184 x 135 x 36 mm (approx 7.2 x 5.3 x 1.4 in) |
Software | ||
---|---|---|
PicoLog for Windows | PicoLog data acquisition software can collect up to 1 million samples. Features include: Multiple views — view data as a graph, spreadsheet or text Parameter scaling — convert raw data into standard engineering units Math functions — use mathematical equations to calculate additional parameters Alarm limits — program an alert if a parameter goes out of a specified range IP networking — transfer measurements via a LAN or over the internet Operating systems supported — Microsoft Windows XP (SP3) to Windows 10. Not Windows RT. |
|
PicoLog languages | Simplified Chinese, Dutch, English (UK), English (US), French, German, Italian, Japanese, Korean, Russian, Spanish | |
PicoLog and PicoSDK | Available from www.picotech.com/downloads | |
Example code | Available from Pico's GitHub organization page, github.com/picotech | |
Optional free software | Drivers for Linux and macOS. |
PC requirements | |
---|---|
Minimum | Processor, Memory, Free disk space: As required by the operating system Operating system: Microsoft Windows 7, 8 or 10, 32-bit and 64-bit versions; macOS 10.9 (Mavericks) or later, 64-bit only; Linux*, 64-bit only Ports: USB 1.1 |
* PicoLog 6 for Linux is distributed as an AppImage, so you can install it without superuser permissions: see appimage.org for further information. The software has been tested on Red Hat, OpenSUSE and Ubuntu.
Documentation languages | |
---|---|
User's guide Programmer’s guide Installation guide |
English, Français, Deutsch, Italiano, Español, 中文 English English, Français, Deutsch, Italiano, Español, Svenska, Nederlandse, Dansk |
General | |
---|---|
Additional hardware (supplied) | USB 2.0 cable, Ethernet cable, user manuals, software CD-ROM |
PC interface | USB 2.0 (USB 1.1 compatible) or Ethernet |
Power requirements | Powered from USB port or Ethernet |
Compliance | European EMC and LVD standards FCC Rules Part 15 Class A RoHS compliant |
Total satisfaction guarantee | In the event that this product does not fully meet your requirements you can return it for an exchange or refund. To claim, the product must be returned in good condition within 14 days. |
Warranty | 5 years |