The TA167 is a dual-range current probe capable of measuring AC and DC current signals up to 2000 A.
Title | Language | Issue | Size | Updated | All products | All languages | All documents | Status |
---|---|---|---|---|---|---|---|---|
TA167 2000 A Current Probe User’s Guide | English | 3 | 276 KB | Mar 28 2017 | english | open | ||
TA167 2000 A Current Probe Frequency Response and Derating Curves | English | 2 | 149 KB | Jun 11 2020 | english | open |
TA167 Current probe specification (all accuracies stated at 23°C ± 1°C) | |
---|---|
Nominal current | 2000 A AC peak or DC |
Measuring range(s) | 200 A / 2000 A |
Output sensitivity | 10 mV/A / 1 mV/A |
Accuracy (0 - 200/1500 A) | ± 1% of reading ± 100 / ± 500 mA |
Accuracy (1500 - 2000 A) | ± 5% of reading |
Resolution | ± 100 / ± 500 mA |
Frequency range | DC to 20 kHz (–1 dB) |
Power supply | 9 V alkaline battery PP3, MN1604 or IEC 6LR61 |
Load impedance (min) | > 100 kΩ and ≤ 100 pF |
Conductor size | 32 mm diameter |
Operating temperature | 0°C to +50°C |
Output cable and connectors | 2 m long coax terminated with safety BNC connector |
Safety standards | EN61010-1:2001 EN61010-2-032:2002 EN61010-031:2002 300 V CAT III, Pollution Degree 2 |