70 MHz 7000 V differential oscilloscope probe 100:1/1000:1

The TA044 is an active differential probe suitable for high common-mode voltage measurement applications up to ±7000 V (DC + peak AC).

This differential probe extends the functionality of standard single-ended input oscilloscopes to allow a safe and accurate method of making high-voltage differential measurements. Applications include making safe measurements in power circuit applications and acquisition of low-speed balanced differential signals found in serial communications buses.

Note: The spring hooks supplied with this probe are rated at 1000 V.

70 MHz 7000 V differential oscilloscope probe 100:1/1000:1 TA044
In stock. Available for despatch.
£759
Title Language Issue Size Updated All products All languages All documents Status
TA044 70 MHz 7000 V Differential Probe User’s Guide English 1 476 KB Nov 07 2016 english open
Choosing the right Pico Technology active differential probe English 4 1 MB Nov 10 2020 english open

Accessories for the 70 MHz 7000 V differential oscilloscope probe 100:1/1000:1

Product Price Quantity
9 V AC power adaptor
PS008
£22
Lemo power cable for Pico TA04x and TA05x active differential probes
TA331
£69
TA044 differential oscilloscope probe specifications
Bandwidth DC to 70 MHz (–3 dB)
Attenuation ratio 100:1/1000:1
Accuracy ±2%
Input impedance 10 M Ω // 10 pF each side to ground
Input voltage 
Differential range* (100:1)
max. differential* (1000:1) 
±700 V (DC + peak AC) or 700 V RMS
±7000 V (DC + peak AC) or 5000 V RMS
Common mode range ±7000V (DC + peak AC) or 2500 V RMS
(100:1 & 1000:1)
Absolute max. voltage (common mode) ±7000 V (DC + peak AC) or 2500 V RMS
(100:1 & 1000:1)
Absolute max. voltage (differential mode) ±7000 V (DC + peak AC) or 5000 V RMS
(100:1 & 1000:1) 
Output voltage
Swing ±7V into 50 kΩ load 
Offset (typical)  <±5 mV 
Noise (typical) 0.9 mV RMS
Source impedance (typical) 50 Ω
CMRR (typical) 80 dB @ 60 Hz; 50 dB @ 1 MHz
Ambient operating temperature –10 to 40 ℃
Ambient storage temperature –30 to 70 ℃
Power requirements 
Standard 4xAA cells or 6 V DC / 200 mA mains adaptor
or regulated 9 V DC / 120 mA mains adaptor
Length of input leads  60 cm
Length of BNC lead 90 cm
Weight 500 g