Posted: Mon Jun 30, 2008 5:44 pm Post subject: Sampling Rates of the 2000 Series Scopes and the ADC Series
The application I am working on requires a source signal at 500Hz pulsed out for 50 ms of each second. I would like to be able to put the time/division setting at 1 sec/div and still be able to see a good representation of the signal. The only way to do this is to sample at4*5 times faster then the source signal. So the question I am trying to find out is what are the sampling rate for the 2000 and ADC series scopes at 1 sec/div (single channel)?
Joined: 07 Jun 2007 Posts: 117 Location: Cambridgeshire
Posted: Tue Jul 01, 2008 12:28 pm Post subject:
Hi,
Our base model 2203 scope can take 2 million samples at 200KS/sec for 10 secs. That is a sample interval of 5uS and is more than enough for your application since it is 400 samples per cycle.
PeterF.