kcosta User

Joined: 17 Apr 2008
Posts: 2
|
Posted: Thu Apr 17, 2008 10:09 am Post subject: Converting ADC 212 Program to PicoScope 2000 Series |
|
|
Hello, I'm trying to convert a program that was originally written for an ADC 212 to use a PicoScope 2202.
In the old program I used the function:
| Code: | | Call adc200_get_single(Values(Int_Channels)) |
to get a single value from the scope.
What is the equivalent function for the ps2000 series?
I have tried:
| Code: | | ps2000_get_values(ps2000_handle, Values_a(0), Values_b(0), values_c(0), values_d(0), overflow, no_of_values) |
with no_of_values set to 1, but it always returns 0.
Any ideas?
Thanks,
Keith |
|
| Back to top |
|
 |
picojohn Site Admin

Joined: 11 Jun 2007
Posts: 391
|
Posted: Thu Apr 17, 2008 1:25 pm Post subject: |
|
|
Hello Keith,
Please try using the PS2000.dll with the programming examples for the PS3000 series, but replacing ps3000 with ps2000 for the routines that are applicable to a PS2000.
Please also refer to the following link: http://www.picotech.com/support/kb/SoftwareInstallation-kb185
Regards
John |
|
| Back to top |
|
 |
kcosta User

Joined: 17 Apr 2008
Posts: 2
|
Posted: Thu Apr 17, 2008 1:42 pm Post subject: |
|
|
Hi John,
Thanks for the reply, I got the example app to work, so I can communicate with the PicoScope 2000, but what I would like to know is which function in the new API can I use that will function as closely as possible to the old one that I used?
The old function was
| Code: | | Call adc200_get_single(Values(Int_Channels)) |
But I can not find a similar function in the new API.
I just need it to return a signal value when it is called. |
|
| Back to top |
|
 |
gruntman Guru

Joined: 28 Sep 2006
Posts: 96
Location: Garden City, KS USA
|
Posted: Thu Apr 17, 2008 3:41 pm Post subject: |
|
|
Keith,
There is not a direct equivalent function for the PS2000 series as the ADC-212's adc200_get_single. The way you would set it up to get a single value in the closest way would be to set the scope up in block mode. First you would need to call the following calls:
1. Open the oscilloscope using ps2000_open_unit().
2. Select channel ranges and AC/DC coupling using ps2000_set_channel().
3. Using ps2000_set_trigger() , set the trigger if required.
4. Using ps2000_get_timebase() , select time bases until you locate the required time interval per sample.
Next you will need to call ps2000_run_block(). On this function you will need to state the number of samples as 1. Then wait until ps2000_ready() is ready. Next call ps2000_get_values() to display the single value. After this you simply have to call ps2000_run_block(), ps2000_ready(), and ps2000_get_values() to get you next single value when you need it. After this is done simply call ps2000_stop().
Regards,
Richard Boyd
Crag Technologies, Inc
http://www.pc-oscilloscopes.com |
|
| Back to top |
|
 |
|